top of page

-86°C Ultra-low Temperature Freezer DW-HL530

M-twindrive -86°C ultra-low temperature freezer built up by dual independent refrigeration systems and the dual cooling systems offers the highest level of protection, if one system unexpectedly fails, the other can maintain the freezer around the -80°C. Valuable and irreplaceable samples can always be securely with the highest confidence even if the compressor fails.

Model

DW-HL530

Cabinet type

Upright

Capacity

530 L

Internal size (W*D*H)

585*696*1266 mm

External size (W*D*H)

930*1041*1947 mm

Net weight

276 kg

Performance

Temperature range

-40~-86°C

Ambient temperature

16~32°C

Climate class

N

Controller

Microprocessor

Display

Touch screen

Refrigeration

Compressor

2

Cooling method

Direct cooling

Defrost mode

Manual

Refrigerant

HC

Construction

External material

Cold rolled steel

Inner material

Stainless steel

Chambers

3 (Stainless steel)

Door lock with key

Yes

Door

1 (2 inner doors)

Access port

2 (Ø 25 mm)

Casters

4 (+2 leveling feet)

Data logging

Built-in USB (Stores up to 1 year of data)

Backup battery

Yes

Alarm

Temperature

High/Low temperature, High ambient temperature

Electrical

Power failure, Low battery

System

Door ajar, Sensor failure, Built-in USB datalogger failure, Communication failure

Accessories

Standard

RS485, Remote alarm contact

Optional

Padlock